| Pulse Oximter Specification Table(A2) | Bitmap
| ||||||||
| Safety(if ever) | |||||||||
| Physical specifications | |||||||||
| Weight(with 2-AAA batties) | 60g | ||||||||
| Weight(without batties) | 35g | ||||||||
| Display | LCD Display | ||||||||
| Size of screen | 29mm×25mm | ||||||||
| Operation Environment | |||||||||
| Temperature: | 5ºC to 40ºC | ||||||||
| Relative Humidity | 30% to 80% | ||||||||
| Atmospheric pressure | 70kpa to 106kpa | ||||||||
| Storage/Transportation | |||||||||
| Temperature: | -20ºC to 60ºC | ||||||||
| Relative Humidity | 10% to 95% (noncondensing) | ||||||||
| Atmospheric pressure | 70kpa to 106kpa | ||||||||
| Measuring specification | |||||||||
| Oxygen Saturation Range(SPO2%) | 70%~99% | ||||||||
| Pulse Rate Range | 30~240BPM(Beat per minuter) | ||||||||
| Measurement Wavelengths(Red) | 660nm | ||||||||
| Measurement Wavelengths(Infrared) | 880 or 904 nm | ||||||||
| Accuracy: | |||||||||
| Blood Oxygen Saturation(SpO2) | ±1%(90% to 99%);±2%(70% to 89%); | ||||||||
| unspecified(<70%) | |||||||||
| Resolution for SpO2 | 1% or 1 digit | ||||||||
| Pulse Rate | 1BPM(Beat per minuter) | ||||||||
| Low Perfusion | ≤0.4% | ||||||||
| Power Supply | |||||||||
| Power Consumption | <20mA | ||||||||
| Power Requirements | Two 1.5 volt AAA-Size batteries | ||||||||
| Battery Life | Approximately 50 hours continues operation | ||||||||
| Alarm Range: | |||||||||
| Blood Oxygen Saturation(SpO2) | <85% | ||||||||
| Pulse Rate | 50-110BPM | ||||||||
| Others | |||||||||
| Automatic Power Off | 8 second no signal | ||||||||
| Alarm | Visual alarm function | ||||||||
新手教學