0100系列氧傳感器;
% range).
Calibration 1. Atmospheric pressure calibration (between 15 to 25% O2) 2. 1000 ppm standard gas calibration (1000±50 ppm)
Analog output 4-20 mA DC isolated, fixed range (0-1000 ppm or 0-25%),
(for 0-25%, 4 mA DC corresponds to 1500 ppm or less).
Contact output Two outputs, rated 250 V AC, 1 A
Hi or Lo alarm, sensor disconnectd alarm
Alarm output High alarm contact output if concentration abnormal
Hi or Lo alarm (depending on specification), setting range : 1 ppm to 24.9%
Self-diagnostics Sensor disconnected
Power supply 100 V AC ±10%, 50/60 Hz, 15 VA
Converter size 96 x 96 x 190 mm
Limiting-current type
Natural diffusion
Temperature 0 - 70°C, Humidity 20 - 80% RH (dew free), Pressure 1013 ± 40 hPa (Abs)
O2 in N2 or mixed gases; cannot be used with silicone vapor, combusible gases or organic solvents
Indoor, panel mounting
0-100 ppm / 0-1000 ppm / 0-10000 ppm / 0-25%, automatic / fixed switching mode
±15 ppm (Meas. range : 20-100 ppm), ±25 ppm (Meas. range : 100-1000 ppm),
±0.3%of reading (Meas. range : 1000-5000 ppm),
±7.5% of reading (Meas.range : 5000-10000 ppm), ±0.25% (Meas.range : 1-25%)
1. Atmospheric pressure calibration (between 15 to 25% O2)
2. 100 ppm calibration (between 80 and 120 ppm)
3. 1000 ppm calibration (between 900 and 1100 ppm)
4. 9000 ppm calibration (between 8500 and 9500 ppm)
5. Atmospheric pressure, 20, 100, 500, 1000 and 9000 ppm calibration
The measuring value can be adjusted to true analyzed value.
1. 4-20 mA DC isolated, fixed range
(0-100, 0-1000, 0-10000 ppm, 0-25% selectable),or 4-20 mA DC autoranging.
2. 1-5 V DC, fixed range (0-100, 0-1000, 0-10000 ppm, 0-25% selectable)
3. Zone signal output for 4-20 mA DC autoranging
Two outputs, rated 110 V AC, 24 V DC, 1 A
Hi or Lo alarm, sensor disconnected alarm
Hi or Lo alarm. setting range : 1 ppm to 24.9%.
Sensor disconnected
Sensor and converter temperature 0 to 70°C (dew free)
100-240V AC ±10%, 50/60Hz 15VA
96 x 96 x 190 mm
Meas. principle
Sampling
Gas sample
Measured gas
Mounting
Meas. rangesunit
Gas outlet cock (with flow adjuster): qty. one
Gas outlet cock (without flow adjuster): qty. one
Standard gas cylinder ( 1000 ppm O2/N2 ) : qty. six
Sensor assy. cable 3 m for OX100 only
Extension cable, 2 m: for OX100 only
Bracket for mounting OX100, two per set.
Sensor assy. One per set : for OX102 only
Extension cable, 4 m: for OX102 only
Extension cable, 9 m: for OX102 only
Extension cable, 29 m: for OX102 only
Sensor holder (metal) for OX100 or OX102
Sensor holder (plastic) for OX100 or OX102
L9811LA plastic resin sensor holder is provided as standard accessory.
*1 For measurement in the range of 1500 to 4500 ppm, use the OX102.
*Standard Product does not have a sensor holder. To get the sensor holder, specify " /S " or " /J".
(If using Sampling Unit, the sensor holder is not required.)
Repeatability
Calibration
Fine tune mode
Analog output
Contact output
Alarm output
Self-diagnostics
Storage environment
Storage environment Sensor and converter temperature 0 to 70°C (dew free)
Power supply
Converter size
Model OX100
Limiting-current type
Natural diffusion
Temperature 0 - 70°C, Humidity 20 - 80% RH (dew free), Pressure 1013 ± 40 hPa (Abs)
O2 in N2 or mixed gases; cannot be used with silicone vapor, combusible gases or organic solvents
Indoor, panel mounting
Meas. principle
Sampling
Gas sample
Measured gas
Mounting
OX100 measuring principles, measured gas, and system configuration are same as for OX102