VOLTAGE | RANGES Frequency Range Peak Continuous Peak < 1 second Input Impedance Display Crest Factor | 900, 215, 46, 10 Vpk 10Hz to 1MHz 1500 Vpk (over voltage) 5000 Vpk (over voltage) 1 MΩ 4.5 Digits 20 (Peak / RMS)
|
RMS 45-65Hz | Accuracy | 0.05% of rdg +0.05% of range |
RMS | Accuracy | 0/1% of rdg + 0.1% of range + 4mV + (0.02 * F)% of rdg |
DC | Accuracy | 0.1% of rdg + 0.4% of range + 5mV |
VOLTAGE +/- PEAK | Accuracy | 0.5% of rdg + 0.5% of range + (0.02 * F)% of rdg
|
CURRENT | RANGES Frequency range Peak Continuous Peak < 1 second
Input resistance Crest Factor | 100, 25, 6.25, 1.6, 0.4, 0.1 Apk 10Hz to 1MHz 20Arms 60Arms (over current) 12.5mΩ 20 (Peak/RMS) |
RMS 45-65Hz | Accuracy | 0.1% of rdg +0.1% of range |
RMS | Accuracy | 0.1% of rdg + 0.1% of range + 1mA + (0.02 * F)% of rdg
|
DC | Accuracy | 0.1% of rdg + 0.4% of range + 1mA |
CURRENT +/- PEAK | Accuracy | 0.5% of rdg + 0.5% of range + (0.02 * F)% of rdg
|
WATTS | RANGES Frequency range | 1W to 90kW 10Hz to 1MHz |
RMS 45-65Hz | Accuracy | [{(Verror/Vreading) +(Aerror/Areading)}*Wreading + 0.1% of W reading |
| Accuracy | 0.2% rdg + 0.1% range + 4mW + ((0.05/PF) * F)% of rdg
|
VA | RANGES Frequency range Accuracy | 1VA to 90kVA 10Hz to 1MHz 0.2% rdg + 0.1 % of range +4mVA + (0.05 * F)% of rdg |
VAr | RANGES Frequency range Accuracy | 1VAr to 90kVAr 10Hz to 1MHz 0.2% rdg + 0.1% range + 4mVAr +((0.05/1-PF) * F)% of rdg |
POWER FACTOR | Power Factor Range Accuracy | -1.000 to +1.000 +/-0.002 +/- ((0.001/PF) * F)
|
FREQUENCY | RANGE Accuracy | DC and 10Hz to 1MHz 0.1% |
VOLTAGE CREST FACTOR | RANGE Accuracy | 1.00 to 20.0 %Vpk error + % Vrms error
|
CURRENT CREST FACTOR | RANGE Accuracy | 1.00 to 20.0 %Apk error + % Arms error
|
PEAK INRUSH CURRENT | RANGE Accuracy | 100Apk 2% of range +/- 20mA |
HARMonIC ANAYLSIS | Number of Voltage & Current Harmonics Maximum Harmonics Frequency Accuracy
Frequency Range
| 50
450kHz 0.2% of Reading + 0.1% of range +0.04% per kHz of Harmonics 10Hz to 450kHz |
THD Total Harmonic Distortion | Range & Accuracy
Formula | Range 0-999% Accuracy 0.4% + (0.1 * F)% of reading Series or difference |
STANDBY POWER | Time Window Resolution | 1-300 sec 1 second |
IMPEDANCE | Range Accuracy | 0.005Ω to 1MΩ 0.2% of Reading +0.1% of range +5mΩ + ((0.05/PF) * F)% of reading |
RESISTANCE | Range Accuracy | 0.005Ω to 1MΩ 0.2% of Reading + 0.1% of range +5mΩ + ((0.05/1-PF) * F)% of reading |
REACTANCE | Range Accuracy | 0.005Ω to 1MΩ 0.2% of reading + 0.1% of range +5mΩ + ((0.05/PF) * F)% of reading |
EXTERNAL SHUNT Scaling | Input Range | +/- 1250 mVpk 0.0001 to 100000 |
MECHANICAL | | 1/2 rack size. Rack height 85mm W = 224mm. Height including feet 103mm D = 285mm. Weight 3.21Kg (7lbs.)
|