Sensitive, economical instrument for environmental testing
? Unique automatic background correction maximizes accuracy ? Detection limits down to 10 parts per trillion (ppt) ? Perfect for EPA method 245.1
This dedicated analyzer enables precise mercury analysis for a variety of applications. Analyze trace Hg levels in water samples and related eco-system impacts. Established water methods ?based on Hatch and Ott cold vapor techniques - include drinking, surface, saline, and waste waters.
The mercury analyzer combines several high-performing instruments to achieve superb detection limits with minimal interferences in a user-friendly platform. Following sample preparation, reduced Hg° vapor is sparged into a long-path 400mm absorption flowcell. User-adjustable parameters include gas flows, lamp currents, and integration times. Absorbance is measured as a function of time and automatically integrated using on-board software. Use up to eight calibration points for mercury quantification.
Background correction is accomplished by acquiring rapidly alternating deuterium light absorbances interspersed with 253.7nm hollow cathode lamp readings. The resultant D2 baseline of non-Hg signals from organic vapors and other interferences is precisely subtracted out to maximize mercury reading accuracy.
Calibration and sample data are displayed on a backlit LCD, with outputs to printer via parallel port, to recorder, and to computer using RS-232. Push the Quick-Purge button to flush residual Hg vapor to an internal trap and ready analyzer for next sample.
What's included: Air dryer with charcoal filter, aerator flask assembly and tubing
SpecificationsSensitivity Detection Limit: 10 ppt
Output RS-232, analog, parallel port
VAC/Hz 230 VAC (50 Hz)
Dimensions 39" W x 12" H x 11" D
Pathlength 400 mm, silica windows
撿出限:10PPt
RS-232輸出
電源:230V(50Hz)
外形尺寸(英寸):39"*12"H*11"D;