Agilent 1200 Series Rapid Resolution LC System - Selected System Specifications |
Pump delay volume | Configurable: Low delay volume configuration: 120 µL Standard delay volume configuration: 600 - 800 µL |
Flow rate range | 0.05 - 5 mL/min |
Maximum pressure | 600 bar |
Column dimensions | Length: 10 - 300 mm, ID: 0.05 - 8 mm |
Temperature range | 10 °C below ambient - 100 °C |
Temperature stability | < ± 0.05 °C |
Max. linear velocity | > 16 mm/sec (water / ACN gradient) |
Injection volumes | 0.1 - 100 µL (1500 µL) without loop change, no overfill (drawn = injected volume) |
Maintenance and system test | Front access to all maintenance parts Maintenance instructions through multimedia CD-ROM Time for full system test (OQ/PV) < 4 hours |
Standard GLP features | Early maintenance feedback-EMF (tracks lamp burn time, usage, number of injections with limits and feedback messages) |
Extended GLP features (for systems inlcuding MWD SL or DAD SL) | Data Recovery Card (DRC) for data security (prevents data losses in case of communication breakdowns) Radio Frequency Identification (RFID) for data traceability. (RFID tags for flow cells and UV lamp allow unambiguous traceability of measurement conditions) |
System control | Through local workstation software, client/server network data system and/or local handheld control module (1200 Series Instant Pilot). Default control: LAN / CAN. |