ng>MEASUREMENT FUNCTIonng> | ng>PA2200Ang> |
Measurement Bandwidth | Range | DC, 0.2Hz to 500kHz |
|
VOLTS Vrmsng> ng> | Display Resolution | 4 3/4 digit |
Measurement Range | 700mVrms to 1000Vrms (1440Vpk) |
(200mVrms with display blanking off) |
Vpeak Ranges | 1440, 720, 220, 110, 33.0, 16.0, 5.00, 2.50 |
Max input volts | 1440Vpk Overload 3000 Vrms for 1 second |
Accuracy | 0.08% Reading +0.1% Range + 10mV |
DC 0.5% Reading +0.5% Range + 10mV |
Common Mode | 90dB |
Input Impedance | 1Mohm 12pF |
|
Current Arms | Display Resolution | 4 3/4 digit |
Measurement Range | 0.01Arms to 20Arms (125Apk) |
Ext Shunt 1mVpk to 4Vpk |
Apeak Ranges | 125, 62.5, 22.0, 11.0, 3.50, 1.80, 0.6, 0.3 |
Max input amps | 20Arms continuous 125Apk |
Accuracy | 0.08% Reading +0.1% Range +2mA |
DC 0.5% Reading + 0.05% Range + 10mA |
Common Mode | 90db |
Input impedance | 0.02 Ohm |
Power active power watts | Accuracy | 0.1% Reading +0.1% Range +2mW |
VA apparent power | Accuracy | 0.1% Reading +0.1% Range +2mVA |
VAR reactive power | Accuracy | 0.1% Reading +0.1% Range +2mVAR |
|
ng>General Specifications:ng> |
Interfaces | Standard | Printer (Centronics) |
Optional | IEEE-488, RS232, Chart Recorder x4 |
Environment | Operating Temperature | 0° to +50° C |
Storage Temperature | -40°C to +60°C |
Humidity | 10% to 90% RH non-condensing |
Dielectric Strength | Inputs to case | 4kV AC 50/60Hz for 1 minute |
AC line input to case | 2kV AC 50/60Hz for 1 minute |
Power | AC line input voltage | 90V to 264V |
AC line input frequency | 45Hz to 450Hz |
AC line input fuse | 20mm 1AT |
Power Consumption | 14W/28VA |
Mechanical | Weight | 5.5kg |
Dimensions (W x H x D) | 215mm x 144mm x 390mm |
|