| ng>Phaseng> | Range | ± 180.00? or 0.00? to +360.00? |
| Accuracy (23? ± 5?C; 1-year cal interval) | 0.1Hz to 99Hz | ± 0.02? ± 1 digit |
| 100Hz to 999Hz | ± 0.03? ± 0.005? /kHz ± 1 digit |
| 1kHz to 99Hz | ± 0.04? ± 0.005? /kHz ± 1 digit |
| 100kHz to 700kHz | ± 0.25? ± 0.003? /kHz ± 1 digit |
| Resolution | 0.01? (0.001? via interfaces) |
| Repeatability | 0.01? ( ±0.003? via interfaces) |
| ng>Frequencyng> | Range | 0.1Hz to 700 kHz |
| Accuracy | ±0.1% (23? ± 5?) |
| ng>Inputsng> | Voltage Range Current ranges using Options 02F/B All voltage and current input combinations are possible | 1mV rms to 350V rms (500V pk) 20mA - 2Arms range Opt. 02F/B 1R00 5Hz - 20kHz 1.0Arms to 20Arms range Opt. 02F/B 0R01 5Hz - 20kHz |
| Bandwidth | 0.01Hz to 700kHz |
| Isolation (inputs to ground) | 2000V |
| Impedance | 2 Mohm and 10pF |
| ng>Environmentng> | Temperature | Operating | 0?C to + 50?C |
| Storage | -40?C to +70?C |
| Humidity (non-condensing) | 10% to 90% RH |
| ng>Dielectric Strengthng> | Inputs to case or AC line input | 2kV AC 50/60Hz 1 minute |
| Power supply to case | 2kV AC 50/60Hz 1 minute |
| ng>Power Requirementsng> | Line input voltage | 110V or 220V AC ± 10% |
| Line input frequency | 50/60Hz |
| Line input fuse | 315mAT |
| Line input conformance | IEC348 Class I; IEC 1000-3-2; IEC801 Parts 1 to 4; EN55011 |
| Power consumption | 25VA |
| ng>Mechanicalng> | Weight | 5kg (6kg in delivery packaging) |
| Dimensions (W x H x D) | 215mm x 144mm x 390mm |