Mach-Zehnder 馬赫-曾德爾幹涉儀
特性
- 兩種型號可供選擇:中心波長為850納米或者1300納米
- 適合掃頻源輸出頻率監測,帶平衡探測輸出
- 低插入損耗
- 平坦的波長響應
- 集成的信號探測,用於功率監測和k-時鐘信號
- 緊湊型設計
Specifications | INT-MZI-850 | INT-MZI-1300 |
Optical | ||
Interferometer Wavelength Range | 780-925 nm | 1225-1375 nm |
Free Spectral Range | 103.3 GHz ± 5% | |
Fiber Type | Nufern 780HP™ | Corning SMF-28™ |
Input/Output Connection | FC/APC Pigtail, 50 cm Long | |
Insertion Loss* | ||
Electrical | ||
Detector Material/Type | Si/PIN | InGaAs/PIN |
Detector Wavelength Range | 320-1000 nm | 800-1700 nm |
Typical Responsivity (max.) | 0.53A/W | 1.0A/W |
MZI Output Bandwidth (3dB) | DC-200 MHz | |
Conversion Gain Power Monitor** | 30 V/W (± 25%) | 60V/W (± 25%) |
Conversion Gain MZI Output** | 30 V/W (± 25%) | 60V/W (± 25%) |
Saturation Power | 100 mW @ 850 nm | 50 mW @ 1300 nm |
Maximum Input Power | 250 mW (Photodiode Damage Threshold) | |
Electrical Output | SMA | |
Impedance | 50 Ω | |
DC-Offset | < ± 5 mV | |
Power Supply | ± 12V, 200 mA | |
General | ||
Size | 120 x 80 x16 mm | |
*包括輸入和輸出尾纖的接頭損耗,在中心波長處測量。 |
工作原理:
迅速掃頻激光光源通常用正弦調諧元件來實現OCT成像應用所需的光學頻率掃描速度。需要準確和可靠的OCT信號重新校準,這樣最終數據點在頻率上才是等間距的。當使用可調諧激光器作為輸入時,Thorlabs的MZI系列幹涉儀時鐘箱組件設計用來產生周期性(正弦)的K時鐘信號。為瞭得到K時鐘信號,正弦信號的最大值和最小值在頻域必須等間隔(見下面的圖1)。兩個最值之間的差值是由自由光譜響應決定的,對於MZI系列該差值是103.3GHz。
|
下面的圖2是該系統的原理圖。內部光纖網絡利用5%的輸入光產生功率監測和K時鐘信號。剩餘的光傳輸到輸出尾纖。特殊設計的耦合器用於為傳輸信號和功率監測信號實現平坦的波長相應(如下面的圖3和圖4)。MZI的兩個信號都是用超低噪聲的放大平衡光電探測器探測的。
|
|
|